Maintenance
3 - 13
ba43121e03
01/2011
3.6
Replenish reagents A + B (C + D)
This routine fills the wet chemical system with the relevant reagents.
The wet chemical system can only be filled with the pairs of reagents,
A and B, or, C and D. The supply and connection of the reagents to the
analyzer module is described in section 1.3.4 C
ONNECTING
THE
PERMEATE
AND
CONSUMABLES
.
Duration
z
approx. 3 minutes for A and B
z
approx. 7 minutes for C and D
Note
You cannot quit the procedure.
At the end of the cleaning procedure, a calibration (AutoCal) is
automatically triggered. This eliminates any possible tolerances of the
consumables.
Starting the procedure
1
Check whether the container of reagents is sufficiently full and
correctly connected (see section 1.3.4 C
ONNECTING
THE
PERMEATE
AND
CONSUMABLES
).
1
Measuring mode
Select the TresCon Pges module with
68
.
2
Call up the
Parameters Pges
menu with
3
. The main menu
appears on the display.
3
Select the
Service
menu item with
68
and confirm with
3
. The service menu appears.
4
Select the
Replenish reagent A + B (C + D)
menu item with
68
and confirm with
3
. A checklist appears on the
display.