Pges
Overview
1 - 15
ba43121e03
01/2011
Note
If the TresCon Pges module is to be connected to an already installed
overflow vessel (determination of several parameters at a single
sampling location), an overflow vessel with 3 permeate nozzles must
be used at this position. If this is not the case, replace the overflow
vessel.
If the TresCon Pges module is to measure a new sampling location, a
further overflow vessel must be mounted.
Connecting the supply
of consumables
Fig. 1-14 Connecting the supply of consumables
Color coding
of the consumables
The tubes of the consumables are nontransparent and color coded:
3
Feed the tubes for the consumables (color coding see below)
out through the holes drilled in the base of the housing to the
outside.
b
lue
y
ello
w
or
ange
g
reen
g
reen
g
reen
g
reen
blue
Cleaning solution (NaOH solution)
yellow
Standard A (dist. H2O)
orange
Standard B
green
Reagent A (peroxydisulfate solution)
green
Reagent B (diluted sulfuric acid)
green
Reagent C (ascorbic acid solution)
green
Reagent D (molybdate solution)