Operation
Pges
2 - 12
ba43121e03
01/2011
2.4
View/change the AutoCal parameters
You can display or change the parameters of the AutoCal function in
this menu. The factory settings correspond to the values obtained from
practical experience in normal wastewater treatment plants. The
Measuring range
parameter is used for information and cannot be
changed.
Note
Displaying the calibration parameters of the last calibration is described
in section, D
ISPLAY
OF
CALIBRATION
DATA
on page 2-10.
Note
The times of AutoCal and AutoClean are linked with one another.
Setting the AutoCal time in this menu also changes the AutoClean time
automatically (see section 2.5 V
IEW
/
CHANGE
THE
A
UTO
C
LEAN
PARAMETERS
on page 2-14).
Parameter
Values
Measuring range
(cannot be
changed)
0.05 to 3.00 mg/L Pges
AutoCal time
Format hh:mm, input step-by-step, minimum
step: 30 min
AutoCal interval
1, 2, 3, 6, 12, 24 h
Standard B
0.01 to 3.00 mg/L PO4-P
(default setting = 1.60 mg/L;
setting the value to < 1.00 mg/L PO4-P does
not make sense)
1
Measuring mode
Select the TresCon Pges module with
68
.
2
Call up the
Parameters Pges
menu with
3
. The main menu
appears on the display.
3
Select the
AutoCal
menu item with
68
and confirm with
3
. The prompt
Presentation of data/Change
appears
.
4
Select the
Change
mode with
68
and confirm with
3
.
5
Enter the PIN digit-by-digit with
68
. Change to the next
digit with
57
and confirm with
3
.
The analyzer module switches to the STOP mode.
Changing parameters