Appendix
Pges
6 - 2
ba43121e03
01/2011
6.2
Pressure sensors and messages
The following table shows the relationship of the fluid level
measurement result and the internal error message, Ex or Hx.
Note
During pressure measurement, the valves are immediately in the mode
before the pressure measurement and the pump is off.
An overview of the relationship of pressure sensors and messages is
given below.
Fluid level
Sample
Dist. H2O
Std. B
NaOH
> 50 cm
Empty (E4)
Empty (E6)
Empty (E8)
4 < h < x cm
OK
OK
OK
-60 < h < 4
cm
Fill
container
(H2)
Fill
container
(H3)
Fill
container
(H4)
< -60 cm
Blocked
(E9)
Blocked
(E3)
Blocked
(E5)
Blocked
(E7)
Fluid level
Reagent A
Reagent B
Reagent C
Reagent D
> 50 cm
Empty
(E91)
Empty
(E93)
Empty
(E95)
Empty
(E97)
4 < h < x cm
OK
OK
OK
OK
-60 < h < 4
cm
Fill
container
(H90)
Fill
container
(H91)
Fill
container
(H92)
Fill
container
(H93)
< -60 cm
Blocked
(E90)
Blocked
(E92)
Blocked
(E94)
Blocked
(E96)
Display on the screen
Type of error
Pressure sensor
Fill reagent container
A (B, C, D)
Reagent con-
tainer A (B, C, D)
must be filled
P
-60 cm < h2 < 4
cm
Reagent A (B, C, D) blocked
Reagent
A (B,
C, D) blocked
P
h2 < -60 cm
Reagent A (B, C, D)