Caution
Push the electrode into the sensor right up to the stop so that the
connection is watertight. Leaks could lead to the destruction of
the sensor.
7
Push the unit into the sensor up to the stop.
8
Screw the protective hood onto the sensor.
9
Carry out the settings for the sensor on the terminal of the
measuring system (see section 3.4).
10
In the case of initial commissioning, condition the sensor toge-
ther with the mounted combination electrode in a diluted stan-
dard solution, e.g. ES/NH4_ISA-10 10 mg/l NH4-N for about 2
hours (see section 6.2).
11
Calibrate the sensor and electrode with the measuring system
(see section 4.1).