AmmoLyt
®
700 IQ
7 - 1
ba75353e05
03/2006
7
What to do if...
No measured value
Measurement does not
function
Measurement provides
implausible measured
values
Cause
Remedy
– Sensor not connected
– Connecting the sensor
– Unknown
– Look in the log book
Cause
Remedy
– Electrode not connected
– Connect electrode
– Liquid has penetrated the
sensor
– Sensor defective, send it back
– Sensor not connected
– Connecting the sensor
– Instrument setting incorrect
– Correct instrument setting
Cause
Remedy
– No calibration performed
– Calibrate
– Electrode not connected or
defective
– Check electrode and electrode
connection
– Electrode contaminated
– Clean electrode
– Liquid has penetrated the
sensor
– Sensor defective, send it back
– Instrument setting incorrect
– Correct the instrument setting
(
Measuring mode
mg/l or mV)
– Potassium compensation is
switched off
– Potassium compensation
with unsuitable value for
potassium content
– Switch on potassium compen-
sation
– Determine and enter potassium
content
– Recalibrate (see also section
4.2.2)