AmmoLyt
®
700 IQ
Calibration and measuring
4 - 5
ba75353e05
03/2006
4.1.4
General course of a calibration
Preparatory activities
An optimum calibration result is possible by,
!
before calibrating,
– conditioning the sensor in the ES/NH4_ISA-10 standard solution
with 10 mg/l NH4-N for approx. 10 minutes
– setting the required calibration procedure (see section 3.4).
!
providing a similar temperature of standard solution and test sample
for calibration
Note
When exchanging an electrode or the entire combination electrode an
initial calibration (see section 4.1.3) always has to be carried out.
Course
Fig. 4-2
Maintenance condition
Note
When the maintenance condition is activated, linked outputs remain in
their current condition. In the measured value display, the measured
value or the condition indicator of the sensor flashes.
Starting the calibration
1
Switch to the measured value display with
m
.
2
Select the measured value display of the required sensor with
d
.
3
Call up the calibration with
c
. The
Maintenance condition:
Linked outputs are frozen.
window appears.
4
Confirm with
g
.
The display appears with a note that an initial calibration may
be necessary.