AmmoLyt
®
700 IQ
Calibration and measuring
4 - 17
ba75353e05
03/2006
Note
Make sure the citation form is correct for all specifications of the con-
centration.
Operating steps during
calibration
Display
Explanation
Cal.: DOUBLE STD. ADD. (5)
Have test sample ready for cali-
bration
Confirm with
g
.
Input sample volume
(100 ... 1000) ml
If necessary select the volume of
the sample with
d
.
Confirm with
g
.
* Rinse electrode.
* Immerse electrode in test sam-
ple.
* Wait for a stable measured
value.
Confirm with
g
. The measure-
ment of the test sample begins.
Cal.: DOUBLE STD. ADD. (5)
Reference voltage of test sample
determined
Have standard ready
Confirm with
g
.
Select standard concentration
(1 / 10) g/l
NH4
If necessary select the concen-
tration of the standard with
d
.
Confirm with
g
.
Add standard to sample
Confirm with
g
.
* Wait for a stable measured
value.
Confirm with
g
.
The measurement begins.