Calibration and measuring
AmmoLyt
®
700 IQ
4 - 12
ba75353e05
03/2006
Operating steps during
calibration
Displays
Explanation
Unscrew protective hood.
Clean and rinse sensor incl. elec-
trode and protective hood, then
reassemble them.
Prepare the sensor as described.
After completing the operating
steps confirm with
g
.
Cal.: 2 POINT STANDARD (3)
Have standard 1 ready for calibra-
tion
Confirm with
g
.
Immerse sensor in standard 1.
Observe minimum immersion
depth (70mm).
Confirm with
g
.
Select standard concentration
(1 / 10 / 100) mg/l
NH4-N
If necessary select another con-
centration of the standard with
d
.
Confirm with
g
.
Subsequently wait for the condi-
tioning time of 15 minutes to
expire. The remaining time is
shown on the display.
Discard used standard. Immerse
sensor in new standard with simi-
lar concentration.
Start calibration.
Confirm with
g
.
Cal.: 2 POINT STANDARD (3)
Calibration values for standard 1
determined
Have standard 2 ready
Confirm with
g
.
Rinse sensor with standard 2.
Immerse sensor in standard 2.
Observe minimum immersion
depth (70mm).
Confirm with
g
.
Select standard concentration
(1 / 10 / 100) mg/l
NH4-N
If necessary select the concen-
tration of the standard with
d
.
Confirm with
g
.
Subsequently wait for the condi-
tioning time of 15 minutes to
expire. The remaining time is
shown on the display.
Discard used standard. Immerse
sensor in new standard with simi-
lar concentration.
Start calibration.
Confirm with
g
.
Содержание AmmoLyt 700 IQ
Страница 1: ...ba75353e05 03 2006 AmmoLyt 700 IQ Operating manual IQ SENSOR NET ammonium sensor...
Страница 10: ...AmmoLyt 700 IQ 2 4 ba75353e05 03 2006...
Страница 20: ...AmmoLyt 700 IQ 3 10 ba75353e05 03 2006...
Страница 44: ...Calibration and measuring AmmoLyt 700 IQ 4 24 ba75353e05 03 2006...
Страница 56: ...Technical data AmmoLyt 700 IQ 8 4 ba75353e05 03 2006...