Calibration and measuring
AmmoLyt
®
700 IQ
4 - 18
ba75353e05
03/2006
4.1.10 Calibration result
Calibration evaluation
After calibrating the system automatically evaluates the calibration data
and current state of the sensor. The drift potential and slope are evalu-
ated separately. For a valid calibration, the values have to be within the
following ranges:
Cal.: DOUBLE STD. ADD. (5)
Reference voltage determined
after first standard addition
Confirm with
g
.
Add standard to sample
Confirm with
g
.
* Wait for a stable measured
value.
Confirm with
g
.
The measurement begins.
Cal.: DOUBLE STD. ADD. (5)
Reference voltage determined
after second standard addition
Confirm with
g
.
Calibration successful
Conc. (NH4-N)
x mg/l
Slope
y mV
Drift voltage
z mV
End of DOUBLE STD ADD (5) cal.
The values for
Conc. (NH4-N)
,
Slope
and
Drift voltage
are dis-
played.
The calibration is complete. Con-
firm with
g
.
The display returns to the
measured value display.
Display
Explanation
Slope:
+50 ... +70 mV
Drift potential:
-45 ... +45 mV