Calibration and measuring
AmmoLyt
®
700 IQ
4 - 2
ba75353e05
03/2006
4.1.2
Overview of the calibration procedures
For ammonium measurements with the AmmoLyt
®
700 IQ sensor, the
following calibration procedures can be selected:
1 point standard (1)
1-point calibration in a standard solution.
1 point ref. (2)
1-point calibration in the test sample. Calibration is carried out determi-
ning the ammonium concentration by an independent reference proce-
dure, e.g. photometry.
2 point stand. (3)
2-point calibration in two standard solutions.
Simple std. add. (4)
Calibration in the test sample with a one-time addition of standard solu-
tion.
Double std. add. (5)
Calibration in the test sample with a double addition of standard solu-
tion.
Determined calibration
data
Depending on the calibration procedure (single-point or two-point), the
following data are determined during calibrating:
In the calibration log and calibration history (section 4.1.10), the slope
and drift potential are output. The drift potential and slope inform about
the age condition of the electrode.
Note
Single-point calibration procedures take over the slope of the last two-
point calibration. If no valid data from a two-point calibration are avai-
lable, the default setting (59.16 mV) is used. In both cases the value is
marked with * in the calibration history.
Calibration procedu-
res
Drift potential:
Slope
1 point standard (1)
"
1 point ref. (2)
"
2 point stand. (3)
"
"
Simple std. add. (4)
"
Double std. add. (5)
"
"
Содержание AmmoLyt 700 IQ
Страница 1: ...ba75353e05 03 2006 AmmoLyt 700 IQ Operating manual IQ SENSOR NET ammonium sensor...
Страница 10: ...AmmoLyt 700 IQ 2 4 ba75353e05 03 2006...
Страница 20: ...AmmoLyt 700 IQ 3 10 ba75353e05 03 2006...
Страница 44: ...Calibration and measuring AmmoLyt 700 IQ 4 24 ba75353e05 03 2006...
Страница 56: ...Technical data AmmoLyt 700 IQ 8 4 ba75353e05 03 2006...