AmmoLyt
®
700 IQ
Calibration and measuring
4 - 13
ba75353e05
03/2006
4.1.8
Calibration procedure,
Simple std. add. (4)
This calibration with a simple standard addition is carried out in the test
sample while adding standard. The volumes of the test sample and
standard have to be dosed exactly.
The following appliances are suitable for dosing:
!
Measuring cylinder to determine the volume of the sample
!
Pipette (microliter pipette as necessary) for accurate dosing of the
standard. The higher the concentration of the standard, the more
important it is to measure the amount of calibration standard very
carefully.
Note
Before
calibrating, determine the potassium content of the sample (see
section 4.2.2).
Before
calibrating, activate the
Potassium compens.
function and
adjust the potassium content (see section 3.4).
The potassium content of the test sample affects the calibration result.
Conditions and
calibration range for
simple standard
addition
Observe the following points for the simple standard addition:
!
The volume of the standard added to the test sample changes the
test sample. Therefore, the volume of the standard should not be
more than approx. 1 % of the volume of the sample.
Example: With 100 ml sample volume, the added volume should not
be more than 1 ml.
!
Through the addition of standard solution, the ammonium content of
the test sample should be increased at least twofold but not more
than tenfold.
Based on these conditions, the range that can be calibrated is from
0.17 to 76 mg/l NH4-N (or 0.22 to 99 mg/l NH4). If the expected measu-
red values are outside this range, a different calibration procedure has
to be selected.
Cal.: 2 POINT STANDARD (3)
Calibration values for standard 2
determined
Confirm with
g
.
Calibration successful
Conc. (NH4-N)
x mg/l
Slope
y mV
Drift voltage
z mV
End of 2 POINT STAND. (3) cal.
The values for
Conc. (NH4-N)
,
Slope
and
Drift voltage
are dis-
played.
The calibration is complete. Con-
firm with
g
.
The display returns to the measu-
red value display.
Displays
Explanation
Содержание AmmoLyt 700 IQ
Страница 1: ...ba75353e05 03 2006 AmmoLyt 700 IQ Operating manual IQ SENSOR NET ammonium sensor...
Страница 10: ...AmmoLyt 700 IQ 2 4 ba75353e05 03 2006...
Страница 20: ...AmmoLyt 700 IQ 3 10 ba75353e05 03 2006...
Страница 44: ...Calibration and measuring AmmoLyt 700 IQ 4 24 ba75353e05 03 2006...
Страница 56: ...Technical data AmmoLyt 700 IQ 8 4 ba75353e05 03 2006...