Maintenance and changing the electrode
The AmmoLyt
®
700 IQ ammonium sensor is maintenance-free.
Warning
Contact with the sample can be dangerous for the user! Depen-
ding on the type of sample, suitable protective measures must be
taken (protective clothing, protective goggles, etc.).
Note
To keep the sensor clean, we strongly recommend to use the CH clea-
ning head (see chapter 6 R
EPLACEMENT
PARTS
AND
ACCESSORIES
).
Note
Please read the maintenance of the combination electrode in the rele-
vant operating manual of the electrode.
5.1
Exchanging the electrode
Note
We do not recommend unscrewing the measuring electrode from the
sensor connection cable when changing the electrode. Otherwise,
moisture and/or dirt can get into the plug connection where it can cause
contact problems.
If you need to disconnect the sensor from the sensor connection cable,
please note the following points:
!
Before disconnecting the sensor from the SACIQ sensor connection
cable, remove any larger pieces of contamination from the sensor,
particularly in the area of the plug connection (brush it off in a bucket
of tap water, wash it off with a hose or wipe it off with a cloth).
!
Unscrew the sensor from the SACIQ sensor connection cable.
!
Always place a protective cap on the plug head of the sensor and on
the SACIQ sensor connection cable so that no moisture or dirt can
get into the contacting surfaces.
Содержание AmmoLyt 700 IQ
Страница 1: ...ba75353e05 03 2006 AmmoLyt 700 IQ Operating manual IQ SENSOR NET ammonium sensor...
Страница 10: ...AmmoLyt 700 IQ 2 4 ba75353e05 03 2006...
Страница 20: ...AmmoLyt 700 IQ 3 10 ba75353e05 03 2006...
Страница 44: ...Calibration and measuring AmmoLyt 700 IQ 4 24 ba75353e05 03 2006...
Страница 56: ...Technical data AmmoLyt 700 IQ 8 4 ba75353e05 03 2006...