Calibration and measuring
AmmoLyt
®
700 IQ
4 - 8
ba75353e05
03/2006
4.1.5
Calibration procedure,
1 point standard (1)
The
1 point standard (1)
1-point calibration is carried out using one
standard solution.
Note
Determine the potassium content of the sample (see section 4.2.2).
Activate the
Potassium compens.
function and adjust the potassium
content (see section 3.4).
The potassium content of the sample does not effect calibration but the
measurement following.
Operating steps during
calibration
Display
Explanation
Cal.: 1 POINT STANDARD (1)
Have standard ready for calibra-
tion
You can use any standard solu-
tion for this.
The ammonium value should be
as close to the value expected for
the test sample as possible.
Confirm with
g
.
Select standard concentration
(1 / 10 / 100) mg/l
NH4-N
If necessary select a standard
with
d
.
Confirm with
g
.
* Rinse electrode
Follow the instructions on the dis-
play. Confirm with
g
.
* Immerse electrode in standard.
* Wait for a stable measured
value.
Confirm with
g
.
As soon as a stable measured
value is reached, the next display
appears.
Calibration successful
Conc. (NH4-N)
x mg/l
Slope
y mV*
Drift voltage
z mV
End of 1 POINT STAND. (1) cal.
The values for
Conc. (NH4-N)
,
Slope
and
Drift voltage
are dis-
played. The slope is taken over
from the last calibration that
determined the slope.
The calibration is complete. Con-
firm with
g
.
The display returns to the measu-
red value display.
Содержание AmmoLyt 700 IQ
Страница 1: ...ba75353e05 03 2006 AmmoLyt 700 IQ Operating manual IQ SENSOR NET ammonium sensor...
Страница 10: ...AmmoLyt 700 IQ 2 4 ba75353e05 03 2006...
Страница 20: ...AmmoLyt 700 IQ 3 10 ba75353e05 03 2006...
Страница 44: ...Calibration and measuring AmmoLyt 700 IQ 4 24 ba75353e05 03 2006...
Страница 56: ...Technical data AmmoLyt 700 IQ 8 4 ba75353e05 03 2006...