AmmoLyt
®
700 IQ
Technical data
8 - 1
ba75353e05
03/2006
8
Technical data
8.1
Measuring properties
Measuring principle
Potentiometric measurement with an ammonium-selective combina-
tion electrode;
Integrated microprocessor electronics, shielded 2-wire connection for
power and data transmission.
Measuring ranges and
resolutions
Temperature
measurement
Temperature
compensation
in the range of 0 °C ... + 40 °C (32 ... 104 °F)
8.2
Operating characteristics
Allowed temperature
range
Allowed pH range of the
measuring medium
4 ... 12 (pH range for measurement: 4 ... 8.5)
Pressure resistance
Sensor with electrode and SACIQ sensor connection cable connected:
Measuring
mode
Measuring range
Resolution
NH
4
+
0.1 ... 129.0 mg/l
1 ... 1290 mg/l
0.1 mg/l
1 mg/l
NH4-N
0.1 ... 100.0 mg/l
1 ... 1000 mg/l
0.1 mg/l
1 mg/l
Voltage
-2000 ... 2000 mV
1 mV
Probe type
integrated NTC
Measuring range
- 5 °C ... + 40 °C (23 ... 104 °F)
Accuracy
± 0.5 K
Resolution
0.1 K
Response time t
99
< 15 s
Measuring medium
0 °C ... + 40 °C (32 ... 104 °F)
Storage/transport
With electrodes mounted:
0 °C ... + 40 °C (32 ... 104 °F)
With electrodes removed:
-5 ... + 65 °C (23 ... 149 °F)
Max. allowed overpressure
2 x 10
5
Pa (0.2 bar)
Содержание AmmoLyt 700 IQ
Страница 1: ...ba75353e05 03 2006 AmmoLyt 700 IQ Operating manual IQ SENSOR NET ammonium sensor...
Страница 10: ...AmmoLyt 700 IQ 2 4 ba75353e05 03 2006...
Страница 20: ...AmmoLyt 700 IQ 3 10 ba75353e05 03 2006...
Страница 44: ...Calibration and measuring AmmoLyt 700 IQ 4 24 ba75353e05 03 2006...