25 | UM-PP-EN-03
3.6.3
North America
Description
Parameters
115 VAC
208/230 VAC
Voltage
Requires at least 115 V from wall outlet.
If voltage falls 10% below 115 V, tread-
mill will shut off and reset.
Requires at least 208/230 V from wall
outlet.
If voltage falls 10% below 208/230 V,
treadmill will shut off and reset.
Wall Outlet Require-
ments
NEMA 5-20 R
(dedicated circuit required)
NEMA 6-20 R
(dedicated circuit required)
Outlet
Compatibility
Standard 3-prong, hospital grade plug
(NEMA 5-20 P)
Will only fit a NEMA 5-20 R outlet
3-prong plug
(NEMA 6-20 P)
Will only fit a NEMA 6-20 R outlet
Fuse
Supply:
20 A Circuit
Device:
Time delay fuse, 1500 A breaking
capacity, 250V, 15 A
Lift fuse
: Time delay fuse, 250 V, 6.3 A
Supply:
20 A Circuit
Device:
Time delay fuse, 1500 A breaking
capacity, 250V, 10 A
Lift fuse:
Time delay fuse, 250 V, 6.3 A
Hospital-Grade Low Leak-
age
For
grounding reliability, only connect to proper receptacle marked “Hospital Grade”
when using for medical use.
Frequency
50/60 Hz
Current
20 Amp
Dedicated line required (cannot share neutral line)