41 | UM-PP-EN-03
5.6.4
Fall Protection with Emergency Stop
The optional fall protection system for the treadmill serves to prevent accidents. It must be used if
there is an increased risk of falling, as falls are associated with a potentially increased risk of injury.
It also increases safety when running on the wide running surface (
28” [
70 cm]). For most users, a
controlled dismount by straddling and standing securely on the side panels is not possible on these
running surfaces.
There is an increased risk of falling (e.g. during performance diagnostics, intense sprinting, and long
runs). There is an increased risk of injury from falling, especially in rehabilitation where patients with
various physical limitations use the treadmill.
5.6.5
Belt Drive Current Limiting
The PPS Series treadmills are equipped with a current limit control function which reduces power
consumption and increases safety. The main safety feature is the current limiter after time overflow.
If the running belt is blocked for more than 10 seconds, the motor current will be reduced to 6A. This
is always recommended in case something gets caught in the belt, as it stops the belt automatically.
Once the current limit control has been triggered, the motor torque is reduced to a minimum to
prevent damage to the motor and electric system.
5.6.6
Dismounting in Emergency Situations
WOODWAY
PPS Series treadmills have a slip-resistant surface alongside the running surface. This
offers additional grip when dismounting and prevents the feet from slipping off of the side panels.
The slip-resistant surface should be checked periodically for wear or lack of grip and replaced if
necessary.
In emergencies, dismount the treadmill as follows:
•
Jump onto and straddle the side panels.
•
The running surface can run between the legs.
•
Then stop the treadmill using the normal STOP button or the emergency stop button.
When a safety chest harness/waist strap is worn, the user can also drop in an emergency if it is not
possible to straddle the running surface.
An alternative is to stand on the side panel with both feet on one side of the running surface, right
or left and to hold on to the railing. This will trigger the emergency stop mechanism via the pull-cord
and the running surface will come to a controlled stop.
WARNING
Components Must Not Interfere With Use of Device!
Adjustment and safety components (e.g. emergency stop pull-cord,
video railing, connected devices) must be secured properly so as not to
interfere with the proper use and movement of the treadmill and user.
Содержание ST PPS 43 Med
Страница 22: ...16 UM PP EN 03 2 7 EC Declaration of Conformity Fig 1 EC Declaration of Conformity...
Страница 26: ...20 UM PP EN 03 3 5 Dimensions Fig 3 Device dimensions PPS Bari Mill 43 Fig 4 Device dimensions PPS Bari Mill 55...
Страница 27: ...21 UM PP EN 03 Fig 5 Device dimensions PPS Bari Mill 70 Fig 6 Device dimensions PPS Med 43...
Страница 114: ...108 UM PP EN 03 12 Maintenance Report DATE MAINTENANCE MEASURES FROM REMARKS...