96 | UM-PP-EN-03
Visual Inspection
According to Section 5.2 of the standard EN 62353, a visual inspection is to be carried out prior to
the measurements. The following points must be checked on PPS Series devices:
1.
Treadmill Operating Manual:
Is the operating manual for the device
immediately available for users and it is valid for the tested treadmill
model?
2.
Accessory Operating Manuals:
Are the operating manuals for acces-
sories and options available?
3.
Labels and Name Plate:
Are all labels on the device legible and com-
plete (name plate, fuse identification, interface labels, labels on the op-
erating and display elements)?
4.
Fuses:
Do the rated values and meltdown characteristics of the inserted
fuses match with the following values: Fuse, operational voltage
250VAC, size 5x20mm (diameter x length), 10A, time-lag (10 AT)?
5.
Visual Condition of the Treadmill
: Is the device undamaged and
properly cleaned? Are slats possibly broken/cracked? Was a visual in-
spection of mechanism (see Section 9.2 Page 91) properly performed?
6.
Use of the Emergency Stop Magnetic Switch:
Is the emergency stop
magnet available with pull-cord and clip, and is this used every time
the treadmill is used in accordance with the manufacturer (determined
by asking the operating personnel)?
7.
Condition of Pull-Cord, Clip and Cord Stopper
: Are the pull-cord of
the emergency stop magnet, the fixing clip, and the cord stopper for
adjusting the length of the cord undamaged and fully functional?
8.
Safe Fall Area:
Is the safety area of 3 ft. W x 6.5 ft. L (1 x 2 m) behind
the treadmill provided?
9.
Power Strips:
Are power strips used? For safety reasons the use of
power strips is forbidden.
10.
Room Circuit Breaker:
Is a line circuit breaker with the following prop-
erties used to protect the line: Rated voltage 250V, rated current 16A
and tripping characteristic C (
“
slow
”
)?
11.
Power Supply:
Are other devices on the same supply line? For safety
reasons the treadmill must be used on a room connection with a sep-
arate line circuit breaker.
Содержание ST PPS 43 Med
Страница 22: ...16 UM PP EN 03 2 7 EC Declaration of Conformity Fig 1 EC Declaration of Conformity...
Страница 26: ...20 UM PP EN 03 3 5 Dimensions Fig 3 Device dimensions PPS Bari Mill 43 Fig 4 Device dimensions PPS Bari Mill 55...
Страница 27: ...21 UM PP EN 03 Fig 5 Device dimensions PPS Bari Mill 70 Fig 6 Device dimensions PPS Med 43...
Страница 114: ...108 UM PP EN 03 12 Maintenance Report DATE MAINTENANCE MEASURES FROM REMARKS...