95 | UM-PP-EN-03
Inspection Intervals
A Technical Safety Check (TSC) must be performed annually by qualified personnel (electricians).
These are "repeated safety checks" in accordance with Section 4.3 of EN 62353:2008.
If the treadmill is rarely used, under certain conditions, the test interval of 18 months may be increased
to a max. of 24 months (see Appendix F of EN 62353). These conditions are as follows:
1.
The device may not be older than 10 years.
2.
The representative must confirm the estimated average weekly use of the treadmill in writing.
3.
The representative must be informed in writing that the TSC inspection intervals must be re-
evaluated when the frequency/intensity of treadmill use increases.
4.
The qualified inspector must consider the environmental conditions and the frequency of past
device malfunctions.