97 | UM-PP-EN-03
Measurements i.a. EN 62353:2008
The values determined in these tests are to be documented together with the measurement method
and evaluated as the basis for comparison for future standards. Measurements are to be carried out
in accordance with Section 5.3 of the standard. The protective conductor resistance (Section 5.3.2),
and device leakage current (Section 5.3.3) are to be measured. A measurement of the leakage current
from the applied part according to Section 5.3.3.3 of the standard is not necessary. For the
measurement of the device leakage current, the direct measurement methods or the difference in
current measurement can be used.
The replacement measuring method may not be used for measuring the device leakage current.
For the measurement of protective conductor resistance, the side panels of the treadmill must be
removed. During the measurement the power cable must be moved over the entire length. If changes
in resistance are observed during movement, it must be assumed that the ground wire is damaged
or has a bad connection.
A measurement of the insulation resistance in accordance with Section 5.3.4 of the standard may NOT
be performed.
If the measured values are between 90% and 100% of the allowable limit, the previously measured
values (reference values) for the evaluation of the electrical safety of the appliance shall be
considered. Note that the measured values of the factory test may differ slightly from the measured
values at each treadmill location due to different test conditions.
The measured values must not exceed the permissible limits specified in the following table:
Measurement
Limit Value
Protective conductor resistance - non-removable power cord
Resistance between the protective conductor of the power cord
and the protective conductor connected, exposed conductive
parts of the unit (treadmill frame and railing)
300 m
Protective conductor resistance - removable power cord
Resistance between the protective conductor of the power cord
and the protective conductor connected, exposed conductive
parts of the unit (treadmill frame + railing)
200 m
Resistance between the protective conductor contacts at each
end of the detachable power cord
100 m
Device leakage current - direct measurement or differential cur-
rent measurement
Measuring procedure defined in Section 5.3.3 of EN 62353
0.5 mA
Function Test
After the examination (inspection and measurement) a functional test must be performed in
accordance with the Function Test found in Section 9.2 on Page 91 of this manual; this is to ensure
that the treadmill has been restored to its necessary condition for intended use, i.e. that it is
operational and safe.
Содержание ST PPS 43 Med
Страница 22: ...16 UM PP EN 03 2 7 EC Declaration of Conformity Fig 1 EC Declaration of Conformity...
Страница 26: ...20 UM PP EN 03 3 5 Dimensions Fig 3 Device dimensions PPS Bari Mill 43 Fig 4 Device dimensions PPS Bari Mill 55...
Страница 27: ...21 UM PP EN 03 Fig 5 Device dimensions PPS Bari Mill 70 Fig 6 Device dimensions PPS Med 43...
Страница 114: ...108 UM PP EN 03 12 Maintenance Report DATE MAINTENANCE MEASURES FROM REMARKS...