![Winncare WINN MOTION 150 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/winncare/winn-motion-150/winn-motion-150_manual_987330021.webp)
L0900
WINNCARE
Oct 2018
580154
- 21 /
58
–
INSPECTION ANNUAL OF PATIENT HOISTS
IDENTIFICATION OF THE MEDICAL EQUIPMENT
ESTABLISHMENT
Category:
Brand/model/type
Serial number:
Inventory number:
Department/Location:
Date of manufacture:
VISUAL INSPECTION
(according to EN ISO 10535)
NA
C
NC
Availability and condition of the instruction manual, loading plate, identification plate/labelling/printing
Good general condition (structure, base, articulations, welds, attachments, etc…). Absence of rusting
Good general cleanliness
Good stability of the hoist and good operation of the base opening/closing
Checking of the tightness of screws (suspension pin on the lifting arm, weighing module attachments,
etc.), absence of play.
Good condition of slings and cradle or hammock, including the attachment system
Visually check that the pin is present in the hole of the safety nut. Visually check that the pin is
depressed approximately 2mm from the outside of the edge of the safety nut.
Good rotation of the gripping bar and good condition of the hooks
Good operation of the wheels and brakes (swivelling, running, locking)
Good lubrication of articulations (absence of noise nuisance, squeaking and hard points, etc.)
Good condition of electric cables, control boxes, connectors and actuators (jacks, motors, etc.)
Good condition of controls and indicator lights
SAFETY INSPECTION
(according to EN ISO 10535)
NA
C
NC
Locking out of operational functions and correct operation of visual and audible alarms
Emergency stop operates correctly
Battery and charger operate correctly
Obstacle stop operates correctly
Correct operation of the emergency descent
Test at maximum load (1 complete cycle )
NA:
Not applicable-
C:
Conform-
NC:
Not Conform
INSPECTION OF THE AMPLITUDE OF MOVEMENTS
NA
C
NC
Satisfactory operation of the jack (elevation/descent)
Maximum and minimum height (according to the manufacturers specifications)
ELECTRICAL SAFETY
(according to EN 60601-1)
NA
C
NC
Tester model:
Serial number:
Continuity :
Breakdown:
COMMENTS:
Dated:
Name:
Inspector’s signature and stamp
Date of the next inspection