9

1. 

The Vacuum Pump must be charged in a well-ventilated area 
between the temperatures of 5 °C (41 °F) and 32 °C (90 °F).  
Charging outside of this range may result in prolonged or 
incomplete charging and may damage your Vacuum Pump.

Only charge the Vacuum Pump in a well ventilated  

area at a temperature between 5 °C (41°F) and 32°C 

(90°F). Charging outside of these conditions may  

damage the Vacuum Pump.

A Vacuum Pump enclosed by a cosmetic cover will 

retain heat, but in most cases should charge normally 

at temperatures below 27°C (80°F). The battery may 

not fully charge at higher temperatures.

2.  Insert the charging plug all the way into the charging port. 

3.  Plug the Charger into a power outlet. The Adapter LED will 

light. If the Node LED does not light, make sure you have 
inserted the charging plug all the way into the port.

 

If the Node LED is red, reset the Charger by unplugging it 
from the wall and plugging it back in.

 

While charging is in progress, the Node LED will be solid. The 
battery indicator on the fob will flash alternately between red 
and green.

4.  When charging is complete, the Node LED will flash.  

Disconnect the Charger from the Vacuum Pump and from  
the power outlet.

When you are ready for the pump to begin regulating the  
vacuum, either turn on the fob (refer to page 11) or press the  

on/off button

 on the Vacuum Pump. The pump will operate  

until the preselected vacuum level is reached.

Note: Your pump can be used while charging. If the pump has 
just been unplugged from the Charger, it will be in Standby State 
(refer to page 11). To resume vacuum regulation, press the on/off 
button one time for Active State. Press again to turn the pump off.

Walking while charging your pump may cause you to trip 

and fall. Do NOT walk while your LimbLogic is plugged into 

the charger.

Use ONLY a WillowWood Charger. Use of any other charger 

may result in damage to the Vacuum Pump or the charger 

or both and may negatively affect EMC performance and/

or compliance and may result in serious injury or death.

The charger may pose both a strangulation and/or  

electrocution hazard and should not be handled by children.

Do not use the Charger near water. The Charger is  

not waterproof. Use of the charger near water may  

 

result in serious injury or death.

Overview

CHARGING THE VACUUM PUMP BATTERY 

The Vacuum Pump should be charged before it is used for the 
first time, at the end of each day of use, and when the Battery 
Indicator on the fob is red.

Node LED 

(indicates Charger function)

RED = fault
SOLID = charging
FLASHING = charged

 Flashing:

Charged

 

 

13.8 0V

 

LLV-21011

solid:

Charging

2012     PN-1988-H

ABN: 22582837497

charging port

Charger 

Vacuum Pump 

To AC outlet

charging plug

Adapter LED

(indicates that 

AC outlet is  

providing power 

to Charger)

Node LED

AC Adapter

Содержание LimbLogic Series

Страница 1: ...Quick Start Guide Quantity 2 for bilateral kit This document provides instruction for the prosthetist in the installation and use of the system A separate document LimbLogic Patient Instructions is i...

Страница 2: ...ventilated area at a temperature between 5 C 41 F and 33 C 90 F Failure to do so may result in prolonged or incomplete charging of the Vacuum Pump and may damage the Vacuum Pump A Vacuum Pump enclose...

Страница 3: ...the following two conditions 1 this device may not cause interference and 2 the user of the device must accept any interference including interference that may cause undesired operation Remarque Le pr...

Страница 4: ...urning On the Vacuum Pump 10 Turning Off the Vacuum Pump 11 Resetting the Vacuum Pump 11 Turning On the Fob 11 Changing Vacuum Set Point 12 Changing Pump State 12 Bilateral Indicators 13 Changing the...

Страница 5: ...n be pried open using a small flat blade screw driver then snapped back together starting at the battery end and working back toward the opposite end The fob battery could cause serious injury or deat...

Страница 6: ...r button until the yellow digits flash in a chasing pattern and then display the current vacuum level The pump is now Active and will operate until the preselected vacuum level has been reached 3 To e...

Страница 7: ...the pump stack the second gasket on top of the first gasket before attaching the pump to the socket You should be able to turn the pump upside down as shown without the gasket falling off If the gask...

Страница 8: ...set at 20 but if those settings are not suitable for your patient you may adjust them as desired The User Maximum cannot be set higher than the system maximum of 20 The optimum vacuum setting is diffe...

Страница 9: ...page 11 or press the on off button on the Vacuum Pump The pump will operate until the preselected vacuum level is reached Note Your pump can be used while charging If the pump has just been unplugged...

Страница 10: ...ton on the Vacuum Pump one time The Vacuum Pump will beep once Before turning on the fob as described below wait 10 seconds for the pump to activate Overview on off button on off button Your battery w...

Страница 11: ...empt to connect to the second pump using the same steps as above If communication cannot be established with either pump the fob will shut down RESETTING THE VACUUM PUMP It is possible for a large sta...

Страница 12: ...le the setting is being applied If the Enter button is not pressed the settings will not be applied and the fob will shut off due to inactivity in 60 seconds CHANGING PUMP STATE Make sure the fob is o...

Страница 13: ...ne of the two white indicators at the top of the fob will illuminate to show which pump left or right is being controlled To switch the fob from controlling one pump to the other press and hold the Up...

Страница 14: ...ximum The User Maximum cannot be set higher than 20 Press the Enter button to apply the new User Maximum setting and advance to the next setting The display will alternate between two quick flashes of...

Страница 15: ...is charging The level of charge is indicated as follows Solid green Vacuum Pump battery is between 50 and 100 charged Flashing green Vacuum Pump battery is between 30 and 50 charged Solid red Vacuum P...

Страница 16: ...nium Pyramid FND 227014 Capscrews Hose Barb connects to external hose on pump VACUUM PLATE PYRAMID The Vacuum Plate LLV 01043 and Vacuum Pyramid LLV 01044 can be used for attaching the Distal Mount pu...

Страница 17: ...t the existing tube just past the Muffler 3 Connect the Exhaust Tube Fitting to the section of tube that does not have the Muffler The pump will function normally but at a slightly higher noise level...

Страница 18: ...ally stalled Refer to the Pump Stall error code information on page 21 Pump battery does not maintain charge for a full day Ensure that the system is properly sealed Ensure that the battery has been c...

Страница 19: ...r at least 30 minutes and try again The pump may have been exposed to a large static charge Press and hold the on off button on the pump until you hear a beep approximately 12 seconds Does the patient...

Страница 20: ...e Fitting from the pump housing Do not lose the o ring from the fitting Install the drain tube and flush the system with alcohol The pump beeps but does not run The Muffler on the Exhaust Tube may be...

Страница 21: ...rature to return to operating range then reset the pump by pressing and holding the pump s on off button for approximately 12 seconds until you hear a beep 03 Motion fail Motion sensor failure Return...

Страница 22: ...um Pump cycles more than once every 20 seconds the Vacuum Pump has a leak If the vacuum level holds steady the pump is functioning correctly 7 Remove the Diagnostic Disc or Sheet and attach the pump a...

Страница 23: ...d abusive activities are defined as activities such as skydiving karate and judo activities that could result in injury to an individual s natural limbs and activities that expose the prosthesis to co...

Страница 24: ...This product complies with applicable standards for sale and use in EC countries This device operates on up to 7 volts of AC and DC current Type BF applied part On Off button Push Push style This prod...

Страница 25: ...e LimbLogic vacuum system has been designed for and clinically tested as a suspension aid It has not been clinically tested for wound healing or other uses WillowWood does not currently support the us...

Страница 26: ...IC 7475A 21011 This device complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 This device must accept an...

Страница 27: ...technicien exp ri ment dans le domaine radio TV ELECTROMAGNETIC COMPATIBILITY Medical electrical equipment needs special precautions regarding electromagnetic compatibility EMC and needs to be install...

Страница 28: ...mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interrup tions and volt age variations on power supp...

Страница 29: ...n elec tromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MH...

Страница 30: ...distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitte...

Страница 31: ...ortation or short term storage between uses Temperature 25 C to 70 C Humidity 15 to 93 non condensing Altitude 1250 to 20000 feet 380 to 6000 meters Pressure 1060 to 470 hPA Long term storage Temperat...

Страница 32: ...Mt Sterling OH 43143 phone 740 869 3377 800 848 4930 fax 740 869 4374 www willowwoodco com PN 2025 Y 28 AUG 2015 Ohio Willow Wood Company B V Keizersgracht 62 64 1015 CS Amsterdam The Netherlands Pate...

Отзывы: