10
TURNING ON THE VACUUM PUMP
CHARGING THE VACUUM PUMP BATTERY IN A VEHICLE
To use a vehicle’s power to charge the Vacuum Pump, you must
purchase a Power Inverter rated at 75 Watts or less to convert the
vehicle’s 12V DC power to 120V AC. An example of an acceptable
Power Inverter is the Digital Concepts™ 75 Watt Portable Power
Inverter (available from general retailers).
1.
Plug the Power Inverter into the vehicle’s 12V DC socket.
2. Connect the Charger to the Vacuum Pump.
3. Plug the Charger’s AC Adapter into the Power Inverter’s
three-pronged receptacle.
Do not operate the Charger above 41˚C (105˚F).
The battery may not fully charge at higher temperatures.
Press the
on/off button
on the Vacuum Pump one time. The
Vacuum Pump will beep once.
Before turning on the fob as described below, wait 10
seconds for the pump to activate.
Overview
on/off button
on/off button
Your battery will power your LimbLogic pump for at least a
full 18 hour day of normal usage. It will do so for the life of the
product. (Refer to “Service Life” on page 23.) If it is unable to
do so, please refer to the troubleshooting section or contact
WillowWood.
CHARGING THE VACUUM PUMP BATTERY OUTSIDE THE
UNITED STATES
Chargers sold for use outside the U.S. come with a set of prong
adapters. These adapters allow you to use the Charger in many
countries.
Plugging a prong set into an outlet without the
Charger may expose you to dangerous voltages
potentially leading to injury or death. Always
plug the prong set into the Charger before plugging
the Charger or the prong set into an outlet.