14
CHANGING THE PROSTHETIST SETTINGS
The system is shipped with the
User Maximum
set at 20 and the
Vacuum Range
set at 4. If these settings are not appropriate for
your patient, you may adjust them as follows:
Make sure the pump and fob are both turned on and are
communicating with each other. Quickly press the
Enter
button
on the fob three times to enable the Prosthetist Settings function.
The three presses must be within approximately two seconds
of each other.
The display will alternate between two quick
flashes of this symbol and the current value for
the
User Maximum
.
Press the
Up
or
Down
buttons to adjust the
User Maximum
.
The digits will flash to indicate that the
User Maximum
is being
changed.
• The user will not be able to increase the
Vacuum Set Point
above the
User Maximum
.
• The
User Maximum
cannot be set higher than 20.
Press the
Enter
button to apply the new
User Maximum
setting
and advance to the next setting.
The display will alternate between two quick
flashes of this symbol and the current value for
the
Vacuum Range
.
Press the
Up
or
Down
buttons to adjust the
Vacuum Range
.
The digits will flash to indicate that the
Vacuum Range
is being
changed.
• The
Vacuum Range
is how much the current vacuum level can
fall below the Vacuum Set Point before the pump activates.
• The
Vacuum Range
cannot be set lower than 4.
Press the
Enter
button to apply the new
Vacuum Range
setting.
The “chasing” pattern will be displayed while the setting is
being applied. The fob will then return to displaying the current
vacuum level.
NOTE: Adjusting the User Maximum and Vacuum Range Settings
can affect the Vacuum Level setting. After changing either the
User Maximum or Vacuum Range settings, please check to
ensure the Vacuum Level is set to the desired level.
Overview