7
Overview
KEYS TO SUCCESSFUL USE
1.
Fabricate a non-porous Total Surface Weight Bearing socket:
• For a transtibial socket, use either a LimbLogic 4-Hole
Attachment Plate or LimbLogic Drop-In Adapter. Follow
the fabrication instructions provided with the adapter.
• For a transfemoral socket, follow the fabrication
instructions provided with the
WillowWood One System
- Socket Fabrication Instructions
booklet.
Prior to delivering the prosthesis, you may want to use the
Diagnostic Disc or Sheet to confirm that there are no leaks
in the system (refer to page 22).
2. Fully charge the Vacuum Pump. (Refer to “Charging the
Vacuum Pump Battery” on page 9.)
3. Properly secure the fully charged Vacuum Pump to the
prosthesis:
•
Press a sealing gasket into the inlet of the pump.
Drawing No.
NA
TJB
XX/XX/XX
0
Part Number
T.J. Braden
DRAWING NUMBER
MAT'L
SCALE
CHK'D BY
DRAWN BY
±0° 30' DEG.
±1/64 FRAC.
±.0005
±.005
±.01
±.04
.xxxx
.xxx
.xx
.x
TOLERANCES, UNLESS OTHERWISE NOTED
PART FINISH
PART NAME
ECO#
BY
DATE
REV
REVISIONS
THE OHIO WILLOW WOOD COMPANY, INC.
Material
"CONFIDENTIAL"
Black Ring Socket.SLDDRW(Default)
Part File: F:\Cheryl Limblogic\Assem1.SLDPRT
PART NUMBER
SHEET-X OF X
Part Finish
MACHINE
Black Ring Socket
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
OHIO WILLOW WOOD COMPANY. ANY REPRODUCTION IN PART OR WHOLE WITHOUT
THE WRITTEN PERMISSION OF OHIO WILLOW WOOD COMPANY IS PROHIBITED.
Be sure to fully seat the gasket onto the inlet before
attaching the pump to the socket.
Failure to properly
seat the gasket may result in leaks.
Note: when using the Drop-In Adapter, two gaskets are
required. After seating the first gasket onto the pump,
stack the second gasket on top of the first gasket
before attaching the pump to the socket.
You should be able to
turn the pump upside
down as shown without
the gasket falling off.
If the gasket falls off,
it was not properly
seated into the recess.
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
OHIO WILLOW WOOD COMPANY. ANY REPRODUCTION IN PART OR WHOLE WITHOUT
THE WRITTEN PERMISSION OF OHIO WILLOW WOOD COMPANY IS PROHIBITED.
Assem1
MACHINE
Part Finish
SHEET-X OF X
PART NUMBER
Part File: F:\Cheryl Limblogic\Assem1.SLDPRT
Assem1.SLDDRW(Default)
"CONFIDENTIAL"
Material
THE OHIO WILLOW WOOD COMPANY, INC.
REVISIONS
REV
DATE
BY
ECO#
PART NAME
PART FINISH
TOLERANCES, UNLESS OTHERWISE NOTED
.x
.xx
.xxx
.xxxx
±.04
±.01
±.005
±.0005
±1/64 FRAC.
±0° 30' DEG.
DRAWN BY
CHK'D BY
SCALE
MAT'L
DRAWING NUMBER
T.J Braden
Part Number
0
XX/XX/XX TJB NA
Drawing No.
Correct
Incorrect
(not seated properly)
THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF
OHIO WILLOW WOOD COMPANY. ANY REPRODUCTION IN PART OR WHOLE WITHOUT
THE WRITTEN PERMISSION OF OHIO WILLOW WOOD COMPANY IS PROHIBITED.
Assem1
MACHINE
Part Finish
SHEET-X OF X
PART NUMBER
Part File: F:\Cheryl Limblogic\Assem1.SLDPRT
Assem1.SLDDRW(Default)
"CONFIDENTIAL"
Material
THE OHIO WILLOW WOOD COMPANY, INC.
REVISIONS
REV
DATE
BY
ECO#
PART NAME
PART FINISH
TOLERANCES, UNLESS OTHERWISE NOTED
.x
.xx
.xxx
.xxxx
±.04
±.01
±.005
±.0005
±1/64 FRAC.
±0° 30' DEG.
DRAWN BY
CHK'D BY
SCALE
MAT'L
DRAWING NUMBER
T.J Braden
Part Number
0
XX/XX/XX TJB NA
Drawing No.
gasket
• For a
Distal Mount
pump: attach the pump and sealing
gasket(s) to the socket, Vacuum Plate (LLV-01043), or
Vacuum Pyramid (LLV-01044) with M6 flathead screws.
For a definitive prosthesis, apply Loctite 242 Removable
Threadlocker (or equivalent) to the screws, and tighten
the screws to 9 ft-lbs (12 Nm).
• For a
Side Mount
pump: attach the pump and sealing
gasket to the Side Mount Lamination Plate on the socket
using the M4x10 socket head cap screws. For a definitive
prosthesis, apply Loctite 242 Removable Threadlocker
(or equivalent) to the screws, and tighten the screws to
22 in-lbs (2.5 Nm).