17
Overview
Muffler
Cut here
Exhaust Tube Fitting
Exhaust Tube Fitting
Exhaust Tube Fitting
EXHAUST TUBE MUFFLER CLEANING
Using a flat blade screwdriver, remove the Exhaust Tube Fitting
from the pump housing. Do not lose the o-ring from the fitting.
Clear the clog in the tube by running water or by gently blowing
compressesd air through the tube.
If the clog cannot be cleared, purchase a replacement Exhaust
Tube from WillowWood (LLV-21080).
If the pump must be used before the replacement tube can be
installed:
1. Remove the existing tube from the Exhaust Tube Fitting.
2. Cut the existing tube just past the Muffler.
3. Connect the Exhaust Tube Fitting to the section of tube that
does not have the Muffler. The pump will function normally,
but at a slightly higher noise level.
POTENTIAL HOME USE CHALLENGES
Your pump, fob, and charger are unlikely to be damaged by
exposure to lint, dust, or light. However, charging the pump
while it is sitting in direct sunlight or near a furnace vent may
heat it sufficiently to prevent it from obtaining a full charge.
While not necessary, outside surfaces can be cleaned by wiping
with a water-dampened cloth.
Do not allow children or pets to play with LimbLogic or its accessories.
While designed for durability, they may be damaged, or configured
improperly if exposed to untrained or rough handling such as chewing.