30
Recommended separation distances between portable and
mobile RF communications equipment and the LimbLogic
LimbLogic is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer
or the user of LimbLogic can help prevent electromagnetic
interference by maintaining a minimum distance between portable
and mobile RF communications equipment (transmitters) and
the LimbLogic as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum
output power
of transmitter W
Separation distance according
to frequency of transmitter m
150 kHz to
80 MHz
80 MHz to
800 MHz
800 MHz to
2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed
above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance
for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by
absorption and reflection from structures, objects
and people.
Technical Information