31
USER PROFILE
Prosthetist user
LimbLogic is intended to be used for assembly of prosthetic
devices by a certified prosthetist or prosthetic technician.
Patient user
LimbLogic is intended for use by patients meeting the
following conditions.
– Age: Any. Determined by clinician’s evaluation of patient
competence and health.
– Weight: (Body weight plus any loads normally or routinely
carried cannot exceed these weight limits.)
<350 lb (160 kg) for U.S. Activity K Level 2 or 3
<300 lb (136 kg) for U.S. Activity K Level 4
– Health: Activity Level K2-K4
– Nationality: Any
– Mentally competent to operate system as judged by
prosthetist.
PERFORMANCE CHARACTERISTICS
Capabilities:
– Vacuum setting up to 68 kPa. For ease of use, this range has
been divided into twenty even steps between 1 and 20.
– Indoor & Outdoor usage
– Normal use
Temperature +5 °C to +50 °C.
Humidity 15% to 93% non condensing
Altitude -1250 to 10000 feet (-380 to +3000 meters)
Pressure 1060 to 700 hPA
– Transportation or short term storage between uses
Temperature -25 °C to +70 °C
Humidity 15% to 93% non condensing
Altitude -1250 to 20000 feet (-380 to +6000 meters)
Pressure 1060 to 470 hPA
– Long term storage
Temperature -25 °C to +25 °C to preserve battery life
Humidity 15% to 93% non condensing
Altitude -1250 to 20000 feet (-380 to +6000 meters)
Pressure 1060 to 470 hPA
– Charging
Temperature +5 °C to +33 °C
Humidity 15% to 93% non condensing
Altitude -1250 to 10000 feet (-380 to +3000 meters)
Pressure 1060 to 700 hPA
Restrictions:
– LimbLogic must not be used in salt water or any other corrosive
environment.
– At temperatures below freezing, condensation will form and
freeze in the pump preventing further operation until warmed.
– The hand-held fob, communicator, and charger should not be
exposed to water.
– Frequency of use: Daily for 16 hours / day or as needed.
PHYSICAL MEASUREMENTS
Module
Dimensions (mm)
Weight (gms)
Fob with Battery
38L x 31W x 8H
17
Distal Mount Pump
105L x 92W x 37H
195
Side Mount Pump
101L x 99W x 36H
200
Charger Wall Transformer 86L x 56W x 37H
261
*Maximum vacuum drawn may be limited at altitudes significantly above sea level.
Technical Information