WillowWood LimbLogic Series Скачать руководство пользователя страница 11

11

Enter 

Current  

vacuum level

Overview

TURNING ON THE FOB

Make sure the pump has been on for at least 10 seconds. 

Press and hold the 

Enter

 (center) button for approximately 

2 seconds until the yellow digits flash in a circular “chasing” 
pattern. The chasing pattern indicates the fob is attempting to 
communicate with the pump. 

When the fob connects to the pump, the current vacuum level 
will be displayed in the center. A connected fob will shut off 
after one minute of non-use. 

If the pump is off or fails to connect to the fob, the chasing  
pattern will end after 6 seconds, and all fob LEDs will turn off. 

For bilateral systems, if the fob cannot establish communications 
with the first pump, the fob will attempt to connect to the 
second pump, using the same steps as above. If communication 
cannot be established with either pump, the fob will shut down.

RESETTING THE VACUUM PUMP

It is possible for a large static discharge to interrupt normal 
operation. Should this occur, press the 

on/off button

 on the 

Vacuum Pump and hold for at least 12 seconds until the unit 
beeps. Resetting the pump may also be required when the fob 
displays certain error codes (refer to page 21).

TURNING OFF THE VACUUM PUMP

Press the 

on/off button

 on the Vacuum Pump and hold for  

1 second. The pump will beep several times to indicate that it 
has been turned off. Refer to the Quick-Start Guide (page 6) for 
more information.

Note: if the pump has just been unplugged from the Charger, it 
will be in Standby State (refer to page 12). To resume vacuum 
regulation, press the on/off button one time for Active State. 
Press again to turn the pump off.

Содержание LimbLogic Series

Страница 1: ...Quick Start Guide Quantity 2 for bilateral kit This document provides instruction for the prosthetist in the installation and use of the system A separate document LimbLogic Patient Instructions is i...

Страница 2: ...ventilated area at a temperature between 5 C 41 F and 33 C 90 F Failure to do so may result in prolonged or incomplete charging of the Vacuum Pump and may damage the Vacuum Pump A Vacuum Pump enclose...

Страница 3: ...the following two conditions 1 this device may not cause interference and 2 the user of the device must accept any interference including interference that may cause undesired operation Remarque Le pr...

Страница 4: ...urning On the Vacuum Pump 10 Turning Off the Vacuum Pump 11 Resetting the Vacuum Pump 11 Turning On the Fob 11 Changing Vacuum Set Point 12 Changing Pump State 12 Bilateral Indicators 13 Changing the...

Страница 5: ...n be pried open using a small flat blade screw driver then snapped back together starting at the battery end and working back toward the opposite end The fob battery could cause serious injury or deat...

Страница 6: ...r button until the yellow digits flash in a chasing pattern and then display the current vacuum level The pump is now Active and will operate until the preselected vacuum level has been reached 3 To e...

Страница 7: ...the pump stack the second gasket on top of the first gasket before attaching the pump to the socket You should be able to turn the pump upside down as shown without the gasket falling off If the gask...

Страница 8: ...set at 20 but if those settings are not suitable for your patient you may adjust them as desired The User Maximum cannot be set higher than the system maximum of 20 The optimum vacuum setting is diffe...

Страница 9: ...page 11 or press the on off button on the Vacuum Pump The pump will operate until the preselected vacuum level is reached Note Your pump can be used while charging If the pump has just been unplugged...

Страница 10: ...ton on the Vacuum Pump one time The Vacuum Pump will beep once Before turning on the fob as described below wait 10 seconds for the pump to activate Overview on off button on off button Your battery w...

Страница 11: ...empt to connect to the second pump using the same steps as above If communication cannot be established with either pump the fob will shut down RESETTING THE VACUUM PUMP It is possible for a large sta...

Страница 12: ...le the setting is being applied If the Enter button is not pressed the settings will not be applied and the fob will shut off due to inactivity in 60 seconds CHANGING PUMP STATE Make sure the fob is o...

Страница 13: ...ne of the two white indicators at the top of the fob will illuminate to show which pump left or right is being controlled To switch the fob from controlling one pump to the other press and hold the Up...

Страница 14: ...ximum The User Maximum cannot be set higher than 20 Press the Enter button to apply the new User Maximum setting and advance to the next setting The display will alternate between two quick flashes of...

Страница 15: ...is charging The level of charge is indicated as follows Solid green Vacuum Pump battery is between 50 and 100 charged Flashing green Vacuum Pump battery is between 30 and 50 charged Solid red Vacuum P...

Страница 16: ...nium Pyramid FND 227014 Capscrews Hose Barb connects to external hose on pump VACUUM PLATE PYRAMID The Vacuum Plate LLV 01043 and Vacuum Pyramid LLV 01044 can be used for attaching the Distal Mount pu...

Страница 17: ...t the existing tube just past the Muffler 3 Connect the Exhaust Tube Fitting to the section of tube that does not have the Muffler The pump will function normally but at a slightly higher noise level...

Страница 18: ...ally stalled Refer to the Pump Stall error code information on page 21 Pump battery does not maintain charge for a full day Ensure that the system is properly sealed Ensure that the battery has been c...

Страница 19: ...r at least 30 minutes and try again The pump may have been exposed to a large static charge Press and hold the on off button on the pump until you hear a beep approximately 12 seconds Does the patient...

Страница 20: ...e Fitting from the pump housing Do not lose the o ring from the fitting Install the drain tube and flush the system with alcohol The pump beeps but does not run The Muffler on the Exhaust Tube may be...

Страница 21: ...rature to return to operating range then reset the pump by pressing and holding the pump s on off button for approximately 12 seconds until you hear a beep 03 Motion fail Motion sensor failure Return...

Страница 22: ...um Pump cycles more than once every 20 seconds the Vacuum Pump has a leak If the vacuum level holds steady the pump is functioning correctly 7 Remove the Diagnostic Disc or Sheet and attach the pump a...

Страница 23: ...d abusive activities are defined as activities such as skydiving karate and judo activities that could result in injury to an individual s natural limbs and activities that expose the prosthesis to co...

Страница 24: ...This product complies with applicable standards for sale and use in EC countries This device operates on up to 7 volts of AC and DC current Type BF applied part On Off button Push Push style This prod...

Страница 25: ...e LimbLogic vacuum system has been designed for and clinically tested as a suspension aid It has not been clinically tested for wound healing or other uses WillowWood does not currently support the us...

Страница 26: ...IC 7475A 21011 This device complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 This device must accept an...

Страница 27: ...technicien exp ri ment dans le domaine radio TV ELECTROMAGNETIC COMPATIBILITY Medical electrical equipment needs special precautions regarding electromagnetic compatibility EMC and needs to be install...

Страница 28: ...mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interrup tions and volt age variations on power supp...

Страница 29: ...n elec tromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MH...

Страница 30: ...distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitte...

Страница 31: ...ortation or short term storage between uses Temperature 25 C to 70 C Humidity 15 to 93 non condensing Altitude 1250 to 20000 feet 380 to 6000 meters Pressure 1060 to 470 hPA Long term storage Temperat...

Страница 32: ...Mt Sterling OH 43143 phone 740 869 3377 800 848 4930 fax 740 869 4374 www willowwoodco com PN 2025 Y 28 AUG 2015 Ohio Willow Wood Company B V Keizersgracht 62 64 1015 CS Amsterdam The Netherlands Pate...

Отзывы: