Catalog No. 29290
26
Rev 3 06/2020
b. Alarm Set Point:
The level at which an alarm activates is called
the alarm set point. On the instrument, the alarm set point is
designated as “
S/P
”. The setpoint is adjustable to any valid
turbidity level over the range of the instrument in steps of 0.01 NTU.
c. Alarm Delay Time:
The alarm delay times are used to prevent
tripping the alarm when the turbidity level is close to the setpoint.
The function of the delay times is as follows:
•
Delay On
: The turbidity level must exceed the alarm set point
continuously for at least this number of seconds before the alarm
activates.
If the “delay on” time is set at 5 seconds and the process
turbidity exceeds the set point continuously for only 4
seconds, the alarm will not be activated. However, if
process turbidity exceeds the set point continuously for 5
seconds or more; the instrument will activate the alarm.
•
Delay Off
: The turbidity level must not exceed the alarm set point
continuously for at least this number of seconds prior to
deactivation of the alarm.
If the “delay off” time is set to 5 seconds and the process has
exited out of the alarm condition, the alarm will be reset only
if the process is out of the alarm condition for a continuous 5
seconds. Otherwise, the instrument will still signal an alarm
condition.
4.2.4.1 Alarm 1 - Configuring Alarm 1
Alarm 1 Function:
The
ALM1
is displayed and the display
indicates the current function of Alarm 1 (
HI
,
LO
,
OFF
or
Error
).
•
Use the
or
buttons to cycle through and select the desired
function.
•
Press the
button to accept it and move to the next menu.
•
If the alarm was turned
OFF
a prompt will appear to set up Alarm 2
(go to section 4.2.10.2). If, on the other hand, one of the other
HI
or
LO
was selected, a prompt will appear to set the delay times.
Alarm 1 Set Point:
Setpoint is indicated by “
S/P
” shown on the
lower row of the display, and is used to select the set point for this
alarm.
•
Select the desired alarm level by using the
and
buttons.