Catalog No. 29290
35
Rev 3 06/2020
4.2.16 Download Logged Files – Select to Download
•
This menu is used to download CSV files to a flash drive. The selections
are No download (NO), Calibration file only (CAL), All Files (ALL) or
individual month (Jan, Feb…). Once the data stick is inserted into the
USB slot on the right side of the instrument push the
button. If no data
stick is inserted you will get an Error (Err). See section 5.4.2
File
Downloads
for more information. The upper display will flash while the
download is taking place and show dOnE when complete.
Setting the data download
: Enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is illuminated, then press
the
button.
•
Press the
button to move to the next menu until
DOWN
is displayed.
•
The selection can be adjusted by using the
and
buttons.
•
Once the selection is made
press the
button to accept it and move to
the next menu or the
button to exit and return to
AUTO.
4.2.17 Date and Time – Setting of Instrument Time & Date
•
The MTOL+ has a displayed date (month / day / year) and time (hour /
minute / Daylight Savings) that must be set upon initial installation/setup.
The next five menus are used for this setup.
Setting Month
: Enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is illuminated, then press the
button.
•
Press the
button to move to the next menu until
MNTH
is displayed,
then press the
button.
•
Change the month by pressing the
or
button.
•
When the desired month has been selected, press the
button to accept
it and move to the next menu or the
button to exit and return to
AUTO
.