Catalog No. 29290
9
Rev 3 06/2020
•
Within 2-3 meters (6 to10 ft) of the sampling point to ensure a quick
response time.
•
Isolated from vibration, and not susceptible to freezing or temperature
extremes.
•
Suitable vertical and flat surface that can support the product’s weight
5 kg (10lbs).
•
20 cm (8 in) free area above the instrument, allowing room for
calibration and cuvette maintenance.
•
The front display should be positioned at eye level or a suitable level to
read and reach controls - approximately 1.4 Meters (4.5 feet) to 2
meters (6.5 feet) from the floor.
•
Confirm that the unit is level.
Figure 3: Mounting Dimensions of the Instrument
Turbidity Sensor
Junction Box