Catalog No. 29290
31
Rev 3 06/2020
To adjust Backlight Brightness
: Enter the
CONFIG
mode by pressing
the
and
buttons until the arrow beside
CONFIG
is illuminated, then
press the
button.
•
Press the
button to move to the next menu until
BRT
is displayed.
•
Change the brightness by pressing the
or
button.
•
When the desired brightness has been selected, press the
button to
accept it and move to the next menu or the
button to exit and return to
AUTO
.
4.2.10 Display Units - Setting Displayed Units
•
The most common unit is
NTU
(Nephelometric Turbidity Units) however
the instrument can display in
FNU
(Formazine Nephelometric Units).
•
All instruments are shipped from the factory set in NTU mode.
Setting/Changing Units
: Enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is illuminated, then press
the
button.
•
Press the
button to move to the next menu until UNIT is displayed.
•
Use the
or
buttons to select the setting
NTU
or
FNU
.
•
Press the
button to accept it and move to the next menu or the
button to exit and return to
AUTO
.
4.2.11 Ultrasonic Cleaning - Cleaning Function
•
This allows for a selection menu to turn off the ultrasonic cleaning
function if desired. The default mode is
On
.
Setting the ultrasonic cleaning
: Enter the
CONFIG
mode by pressing
the
and
buttons until the arrow beside
CONFIG
is illuminated, then
press the
button.