Catalog No. 29290
28
Rev 3 06/2020
4.2.5 Offset – Offset Reading Adjustment
•
A reading offset can be made to allow for the MTOL+ to agree with
another instrument if desired.
•
This range is limited to ± 1NTU.
•
For more information refer to section 4.3.6.
To change the Offset:
Enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is illuminated, then press the
button.
•
Press the
button to move to the next menu until
OFST
is displayed.
•
Select the offset using the
or
buttons to turn the function
On
.
When the OFST is turned ON it will stay illuminated in
AUTO as a reminder that it is being used.
•
Select the offset using the
or
buttons.
•
Press the
button to accept the setting and move to the next menu or the
button to exit and return to
AUTO
.
4.2.6 Access Code - Setting Security
•
The MTOL+ is equipped with a security access feature that can be
activated. If this option is turned ON, the user is required to input an
access code into the instrument to get to any mode other than
AUTO
. A
three-digit code is used.
•
HF Scientific highly recommends the use of a security access code
in order to help prevent unauthorized access to the system and data.
When the
CODE
is turned
On
the icon will be illuminated in
AUTO
as a reminder that an access code is being used.
Setting Security on/off:
Enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is illuminated, then press the
button.
•
Press the
button to move to the next menu until
CODE
is displayed.
•
Use the
or
buttons to select the setting. (
On
or
OFF
).
Содержание HF scientific 28052