Catalog No. 29290
52
Rev 3 06/2020
The cuvette can be replaced by first shutting off the flow using the
provided shutoff clamp; unscrewing the old cuvette and replacing with a
fresh clean one.
5.8 Replacing the Source Lamp
The source lamps in the MTOL+ instruments are designed for long life.
The IR lamp and white light version have long lives. These lamps,
however, are not covered by the warranty. If the lamp should need
replacement, we recommend calling HF Service Department for
assistance.
5.9 Optional Factory Installed 24V Option
This factory installed option allows for the operation of the MTOL+ from 24
volts DC. For this instrument, the changes are in the junction box only.
The range of correct operation is from 18-36VDC. Caution must be used
to not exceed this voltage.
Connections are not polarity sensitive and are made to the same place as
mentioned in section 3.4.1.
Order numbers are 28052-24V for the IR version and 28053-24V for the
White Light version.
All other specifications remain unchanged.