Catalog No. 29290
47
Rev 3 06/2020
5.3 Routine Measurement: AUTO Mode
After the instrument has warmed up and the instrument is operating in
AUTO mode, readings will be updated (posted) once per second. These
updates occur to the screen, Modbus, 4-20mA and can be logged
internally to the instrument. These updates also affect the operation of the
alarms. Please note that these updates are all affected by the setting of
the Signal Averaging (see section 4.2.7) and possibly Offset (see section
4.2.5).
If power is lost at any time, once power is restored, the instrument will
restart and the instrument will continue in AUTO operation with the same
settings. There may be some slight instability if the instrument cooled off.
If the onboard data logging was used it will record 0.00NTU during the
period that instrument is not running.
The MTOL+ can read up to 110% of the range. Readings above 100%
of the range may not be accurate. After 110% of that range the
display will flash indicating an over-range condition. The 10 NTU
range can display up to 11 NTU, the 100 NTU range can display up to
110 NTU, and the 1000 NTU range can display up to 1100 NTU.
5.4 Data Retrieval
Data is available in several forms. It can be viewed on the screen, logged
in the instrument, read via Modbus or output the 4-20 mA.
5.4.1 USB Connection
The USB connection can be used to upload new software or download
logged data and calibration reports via a USB flash drive.
Software uploads can be made by loading the software onto a USB flash
drive and inserting it into the USB connector. The update will not affect
configurations or stored data, however it always advisable to check the
configuration as menus may change. Check our website:
for software upgrades. The instructions are in
section 4.2.18 of this manual.
5.4.2 File Downloads
•
For instructions see 4.2.16
•
All files download in a CSV (comma separated value) file format. These
are easily imported directly into most spreadsheet and database
programs.
•
The main data includes the Date, Time, Reading, Units, Alarm 1 and
Alarm 2. The instrument ID is part of the file name.