6.3 Bubbles in the Sample
If bubbles are present in the sample water the readings may appear
higher than expected. These reading may also not be stable.
Remove cuvette if air can be seen coming into the flow-through unit from
the inlet tube, this is air introduced, probably from a small water or air leak.
This will need to be resolved in the plumbing.
If air cannot be seen entering the intake tube but appears on the inside of
the glass cuvette, it is coming out of solution. This can usually be cured by
applying pressure to the rotary backpressure clamp on the OUT side of
the cuvette. This helps keep the air in solution so it will not be seen by the
instrument. Refer to figure 4.
6.4 Technical and Customer Assistance
If for any reason assistance is needed regarding this instrument please do
not hesitate to contact either the HF scientific Technical Service
Department or the HF scientific Customer Service Department:
HF scientific
16260 Airport Park Drive, Suite 140
Fort Myers, Florida 33913
Phone: (239) 337-2116
Fax: (239) 454-0694