9.1.3
FDX3543RPW model
1. Connect the power supply to the connector placed on
the back of the Battery Charger
2. Insert the discharged battery in the Battery Charger.
Ensure that the battery insertion makes a click.
A - Terminal of Battery Charger
B - Knobs to remove the battery from Battery Charger
C - Signal Led:
OFF: disconnected from the mains or not in charge
red: fast recharge mode (up to 90%)
green: standard recharge (from 90% to100%)
blue: error
In case during the battery charge the signal led
is blue, stop the charge and check the terminals of the
Battery Charger.
If the error persists, stop immediately the battery
charge.
9.1.4
Battery Charger
The Battery Charger is not a medical device and it mustn't be placed nearby the patient.
The dimensions in the figure show the min. extension
of the patient environment in a surrounding area that is
hurdles-free.