8.6
Radiological data
Description
Data
Nominal power (IEC 60601-1)
32kW @100kV, 320mA, 100ms
kV values
40 ÷ 125kV at steps of 1kV
kV accuracy
±5% (IEC 60601-2-54)
mA values @115/230Vac
50 ÷ 400 mA
mA accuracy @115/230Vac
±10% (IEC 60601-2-54)
mAs values @115/230Vac
0,1 ÷ 220 mAs
mAs accuracy
±10% (IEC 60601-2-54)
Exposure times @115/230Vac
0,001 ÷ 2,2 s according to mAs
Time accuracy
±10% (IEC 60601-2-54)
In DR the mAs and the exposure times are limited respectively at 110 mAs and 1s by the FPD panel
features.
8.6.1
kV-mAs relationship
from kV
to kV
mAs (*)
mAs (**)
---
40
0.1 ÷ 110
0.1 ÷ 220
41
45
0.1 ÷ 100
0.1 ÷ 200
46
52
0.1 ÷ 90
0.1 ÷ 180
53
62
0.1 ÷ 80
0.1 ÷ 160
63
72
0.1 ÷ 71
0.1 ÷ 140
73
92
0.1 ÷ 56
0.1 ÷ 110
93
112
0.1 ÷ 50
0.1 ÷ 100
113
125
0.1 ÷ 45
0.1 ÷ 90
(*) in DR; in CR/Film 1 capacitors group
(**) in CR/Film 2 capacitors groups