7.9
Optional: radiography with examination table or Potter Bucky grid
The examination table or the Potter Bucky must be in compliance with IEC 60601-1 standard and
must be connected to the equipment according to 93/42/CE medical devices directive and following
revisions.
In LOCAL mode: fully extend the extensible cable of the x-ray control and stay as far as possible
from the radiation source.
In REMOTE mode: stay as far as possible from the radiation source.
The mode "Radiography with examination table or Potter Bucky grid" is active in CR mode or with
Digital Imaging System OFF.
1. Place the cassette and the patient according to
requirements.
2. Connect the examination table or the Potter Bucky
grid to the connector (A) placed on the front part of the
equipment.
If the connection with the examination table or with the
Potter is right, the display shows the relative icon.
3. Press the (A) push-button to activate the Potter
Bucky.
4. Place the monobloc and collimator, set the exposure field.
5. Select kV and mAs values in manual or APR mode by selecting the data concerning the exam to be
performed.
6. Hold the wired x-ray emission control or the remote control (optional)
7. Go as far as possible from the x-ray source.
8. Perform the exposure. At the end of the exposure the acoustic signal of the equipment emits three beeps.
When the examination table or the Potter Bucky grid have been selected, one of the most frequent
difficulties is the "no consent to proceed" of the Potter Bucky grid: in this case check the
connection.