Error messages
Typ.
Text
Meaning
Intervention
F
FAULTY POWER
Error on the Charger or Chopper.
Energy not available
Switch OFF, wait some minutes,
switch ON, if the error appears gain
call Service
F
ERROR V3
Absent V3 power supply
Switch OFF, wait some minutes,
switch ON, if the error appears gain
call Service
F
FILAMENT
Filament current not present
Switch OFF, wait some minutes,
switch ON, if the error appears gain
call Service
F
ERROR V2
Power supply V2 not present in the
circuit of mA and kV set
Switch OFF, wait some minutes,
switch ON, if the error appears gain
call Service
F
LOCKED
STARTER
Error during the start time
Press the RESET key to go on, repeat
x-rays
F
LACK OF X-RAY
The kV have not reached 75% of the
value set within the first 10mS of
exposure or lackage of kV
Press the RESET key to go on, repeat
x-rays
F
MAX. TIME
The max. exposure time has been
reached.
Press the RESET key to go on, repeat
x-rays
F
DATA ERROR
Memory error, data checksum error
Switch the equipment OFF, wait
some minutes and then switch it ON
again, if the error appears again call
Service.
F
INVERTER KV
ERROR
During the XR emission the kV are
decreased under 75% or increased
over 110% of the value set or the
high voltage circuit unbalanced
during the exposure
Press the RESET key and repeat the
exposure
F
INV. OVERLOAD
Power to the inverter out of scale
Press the RESET key and go on
F
INVERTER ERROR
IGBT drivers error
Press the RESET key and go on
F
TUBE CALIB. ERR.
Error of XR tube calibration
Call Service.
F
XR HANDSWITCH
ERR.
Faulty x-ray handswitch
Check the integrity of the x-ray
handswitch, switch OFF and switch
ON the equipment, try again, if the
error persists call Service