Visitor T30 C-DR
User's Manual
[File: 218899-01-01-B.docx]
Page 49/74
8
TECHNICAL SPECIFICATIONS
8.1
Electrical data
Description
Data
Power supply
115 ÷ 230Vac ±10%,
standard monophase with ground conductor.
Automatic setting of the equipment according to
mains voltage
Frequency
50/60 Hz ± 5 Hz
Absorbed current
10 A
Line compensation
Automatic
Line resistance
<1 Ω @ 115/230Vac
Standard outlet
16 A @ 230Vac
Power supply cable
8 m
Insulation class
Class I with applied parts type B
Use conditions
Continuous working with intermittent load
Classification according to the liquids seepage
IPx0
Safety in presence of anaesthetic inflammable gases
The equipment is not type AP or APG
8.2
Environmental conditions
Environmental Factor
In normal use
Warehouse and transport
Temperature
from 10 °C to 40 °C
from -25°C to 70°C
Relative Humidity
from 30 % to 75 % non-
condensing
from 10% to 90% non-condensing
Pressure
from 700 hPa to 1060 hPa
from 500 hPa to 1060 hPa
8.3
Total filtration of the equipment
Description
Data
Monobloc
1,1 mmAl @75kV
Additional fixed filter
0
Collimator
2 mmAl @75kV
Total filtration of monobloc group
3.1 mmAL @75kV
Additional filtration of DAPMeter
0.3 mmAl @75kV
Total filtration
3.4 mmAl @75kV