2.2
Electrical Safety
This equipment is in compliance with safety class I, Type B, in accordance with IEC 60601-1 norm.
Do not use the equipment near or leaned against other equipments.
Do not remove protections or cables from this equipment, unless it is expressly required in this
User's Manual, because inside it there are dangerous electrical voltages. The removal of protections
or cables can cause mortal injuries or serious damages to the people.
Protections or cables must be removed only by qualified and authorized technical personnel. Use the
equipment only in rooms or areas comply with all applicable laws ( or regulations having the force of law),
referring to electrical safety of this type of medical device.
Always insulate the equipment from the power supply before proceeding with cleaning or disinfection
operations in order to avoid electric shocks.
Equipotential earth connection
The equipment is supplied with an equipotential earth connection point.
The equipment can be used only in areas comply with local electrical safety norms and in environments
suitable for medical activities. Besides IEC 60601-1 norm provides instructions about the equipotential earth
connection point.
Additional equipotential earth connection
An additional equipotential earth connection is provided because the equipment is movable and the reliability
of the main equipotential earth connection point can be insufficient.
It is possible to use this equipment only in rooms comply with IEC norm requirements.
The equipment described isn't protected against liquids seepage. Its classification is IPx0.
2.3
Mechanical safety
Be sure that parts of the body or clothes aren't stuck among moving components of the equipment.
Remove all objects from range of motion of the equipment.
Check that the unused hanging components ( monitor and radiogenic complex) are positioned so as
not to affect neither the user nor the patients.
It is not possible to transport this equipment while it is working. For a safety transport, switch off
the equipment before transporting it and ensure that all system peripherals ( monitor, mouse,
keyboard, cables etc.) are disconnected.
Do not remove protections or cables from this equipment, unless this operation is expressly
requested in this User's Manual.
The equipment includes moving parts. The removal of protections can cause serious or mortal
physical injuries to people.