2.10
FPD,warnings and precautions
Risk of damages to people
Anomalous use of the FPD component.
► Don't use the component for uses other than those foreseen.
► Don't use the component as main shield against x-rays.
Risk of damages to FPD component
The FPD component has internally some shock sensors that allow to know when the shock is
occurred and the type.
If the shock exceeds precise values, the component must be readjusted or returned to factory for the
repair if the shock level is excessive.
► Don't let fall the FPD component.
► Handle with care.
Risk of damages to FPD component
The FPD component is not a patient support device.
It must be placed on a suitable surface, such as a table or the floor, before applying the patient
weight to the component.
► Observe the limitations on the max. weight concentrated on a small area of the FPD surface.
► Observe the limitations on the max. weight divided uniformly on the whole FPD surface.
The limitations on the weight are indicated in the technical specifications of every single FPD panel.
Hot surfaces.
During the standard use the surfaces of the FPD component become hot because of the dissipation
of the component power.
The temperature increases according to the image acquisition frequency.
Inside the component there are some temperature sensors whose value is indicated in DIS user
interface.
If the internal temperature of the FPD panel exceeds the value for every specific panel, switch OFF
and let cool the component.
► Don't use the FPD component to supply heat to the patient.
► Don't exceed the temperature limits indicated for every single FPD component.
► Don't leave the FPD component near heating sources.
Before using the FPD component, read carefully the relative User's Manual attached to the
equipment.