![Villa Visitor T30 C-DR Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/villa/visitor-t30-c-dr/visitor-t30-c-dr_user-manual_1026484043.webp)
Visitor T30 C-DR
User's Manual
[File: 218899-01-01-B.docx]
Page 43/74
7.8
Operation mode
The equipment manages two exposure operation modes that can be selected by the operator at any minute:
LOCAL: the exposure is performed with the wire control with extensible cable supplied with the equipment.
REMOTE: the exposure is performed with the infra-red remote control (accessory), if installed.
At the start up the mode is LOCAL.
1. Press the key (A) to change exposure operation
mode. The led ON near the key indicates the REMOTE
mode.
An exposure mode excludes the other one.
LOCAL mode
Ionizing radiations.
The operations described below require the emission of ionizing radiations.
► Take the proper measures in order to avoid exposing any part of the body to direct or indirect
radiations.
Completely extend the extensible cable of the x-ray control and keep at least two meters from the
radiation source.
The cable of the x-ray control can be extended without irreversible deformations till 4 mt.
Beyond 4 mt the deformation becomes irreversible and the x-ray control must be replaced.
The x-ray emission control consists of a double step
push-button:
1° step : preparation
2° step : exposure control
Hold down the x-ray emission control as long as the
exposure expires, then release it.
When the exposure has been completed, the equipment
emits three beeps as signal.
It is possible to fully press the x-ray emission control ("exp" position) from the beginning. In this
case the x-ray exposure will be automatically performed after the preparation.