![Villa Visitor T30 C-DR Скачать руководство пользователя страница 41](http://html1.mh-extra.com/html/villa/visitor-t30-c-dr/visitor-t30-c-dr_user-manual_1026484041.webp)
Visitor T30 C-DR
User's Manual
[File: 218899-01-01-B.docx]
Page 41/74
7.7.4
APR data table
APR programs suggested and preloaded by the manufacturer on the equipment, only represent
recommendations to be applied to the patient in order to optimize the operation and result of the
examination.
kV and mAs values set at the factory in APR programs can be modified at any moment by the user.
They can be stored only if, during the equipment configuration, this possibility has been set (by
authorized personnel only).
In the following table the dose values are expressed in µGym
2
(DAP Dose Area Product), the exposures are
set with standard values for clinical investigations on the patient and performed with the x-ray tube with a SID
(Source - Image receptor Distance) of 100 cm without grid.
The dose measurement has been performed according to IEC 60601-1-3 § 5.2.4.2 with dosimetric chamber
supplied with the equipment and collimator with all shutters opened.
"BODY" folder
Name
Focus
kV
mAs
µGym²
Thorax AP
LF
110
2
17.83
Abdomen AP
LF
81
16
76.37
Abdomen LAT
LF
90
20
118.94
Thorax LAT
LF
110
4
34.63
Breastbone
SF
85
4
22.84
Ribs
SF
70
6,3
24.12
"CRANIUM/vertebrae" folder
Name
Focus
kV
mAs
µGym²
Cranium AP
SF
77
10
46.71
Vert. dors. AP
SF
77
20
93.97
Vert. lumb. AP
LF
81
20
104.10
Vert. cervic.
SF
66
8
27.07
Vert. dor. LAT
LF
81
16
76.39
Vert. lumb. LAT
LF
90
20
118.84
"UPPER EXTREMITY" folder
Name
Focus
kV
mAs
µGym²
Clavicle
SF
66
5
16.81
Humerus
SF
60
3,2
8.56
Elbow
SF
55
4
8.53
Forearm
SF
55
2
4.27
Wrist
SF
50
2
3.25
Hand/Fingers
SF
46
1,6
1.99