![Villa Visitor T30 C-DR Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/villa/visitor-t30-c-dr/visitor-t30-c-dr_user-manual_1026484065.webp)
Visitor T30 C-DR
User's Manual
[File: 218899-01-01-B.docx]
Page 65/74
8.13
Optional: DAP Meter
Description
Data
Model
Diamentor CI-P
Type
Device for the area-dose product measurement in x-
ray diagnostics according to IEC 60580 standard.
Principle of measurement
Radiation measure with ionization chamber
Measured quantity
Area-dose product
Digital resolution
0,01 μGym2
Max linearity error
< 2.5%
Nominal range of dose-area product rate
(0,01 ÷ 2500) μGym2 /s
Equivalent filtration of the chamber @75kV
0.3 mm Al
Max measurement field
118 x 118mm
Dimensions (W x D x H)
152 x 234 x 23 mm
Weight
455g