![Villa Visitor T30 C-DR Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/villa/visitor-t30-c-dr/visitor-t30-c-dr_user-manual_1026484044.webp)
User's Manual
Visitor T30 C-DR
Page 44/74
[File: 218899-01-01-B.docx]
REMOTE mode (optional)
Ionizing radiations.
The operations described below require the emission of ionizing radiations.
► Take the proper measures in order to avoid exposing any part of the body to direct or indirect
radiations.
Keep as far as possible from the radiation source.
The wireless remote control works is visual mode up to 10 mt with an operating angle of >180°,
event through glass and leaded glass.
It doesn't work through doors or walls.
Activation
1. Remove the remote control IR from its support base,
move away as much as possible from the x-ray source
and address the remote control towards the receiver on
the equipment.
2. Press the key “LAMP” (A) to turn the collimator
lamp ON.
3. Press and release the key “RAD” (B) to start the
exposure preparation phase.
4. Press and hold down the key “RAD” to perform the
exposure.
5. After performing the exposure, put again the remote
control in the support base on the equipment.
Deactivation
The light of the collimator deactivates automatically by pressing the key “RAD” or by pushing the key
“LAMP”.
The preparation cycle stops automatically and the stand-by condition returns if the exposure is not controlled
within 15sec from its activation.
At the end of the exposure the equipment returns in stand-by even if the key “RAD” is held down. In order to
perform a new exposure, it is necessary to release the key.
The exposure in progress can be stopped by releasing the key “RAD” ("
dead man”
control).
In case the receiver should lose the received signal because of insufficient intensity or interference of a body
or an object that inhibits the IR signal to arrive to the receiver, therefore the active controls are immediately
removed. For performing a new control, it is necessary to release the key and then press it again.
Remote anti-loss research
In case the remote control has not been put into its support base for 2 minutes from the last press of a key, an
audible signal, a series of two "beeps" every 5 seconds, is emitted.
In order to stop the audible signal, put the remote control again in its support base, or press one of the two
keys. This press does not imply any control, in order to perform a control it is necessary to press it twice.
When the remote control is in its support base, it switches OFF and there is no batteries consumption any
more. In this condition no control is released by pressing the keys.