Page 45/74
Indication of discharged batteries
The indicator of discharged batteries (red LED) flashes with 1 flash per second when the batteries have a
limited autonomy (recommended replacement) and with 2 close flashes every second when the batteries are
discharged (mandatory replacement).
7.8.1
After the exposure
An acoustic signal (three beeps) indicates the correct
exposure progress.
The performed exposure time is displayed in ms.
The writing “BUSY” replaces “READY” till the
equipment is ready for the following exposure.
The writing “Exposure done” flashes for about 15s
between the kV and mAs data.
If the x-ray button is held down at "1" click (preparation phase) for more than fifteen seconds
without pressing the "2" click (X-ray releasing phase), the message "OVERTIME" is displayed.
Release the pushbutton, press the key RESET and repeat the exposure.
If the x-ray emission control is released before the end of the exposure, the message "X-RAY
MANUAL STOP" is displayed and the exposure is interrupted.
The radiological data effectively released are displayed.
Press the key "RESET" and repeat the exposure.
7.8.2
DeLuxe
Once the exposure has been completed, the preview of
the image processed with the acquisition parameters is
displayed.
It is possible to accept the image, reject it and perform
another exposure.
After acquiring all images, it is possible to send the
images to the hospital system or print the copies.
Refer to the operating manual enclosed to the equipment for further information regarding the
Digital Imaging System use.