3.11.10 Error messages
Description
Error message
Protocol
Parity error
[E:][000001][CR][LF]
Framing error (data length, number of stop bits)
[E:][000003][CR][LF]
Input buffer overflow (to many characters)
[E:][000002][CR][LF]
Commands
<CR> or <LF> missing
[E:][000010][CR][LF]
: missing
[E:][000011][CR][LF]
Unknown command
[E:][000020][CR][LF]
[E:][000021][CR][LF]
Invalid value
[E:][000022][CR][LF]
[E:][000023][CR][LF]
Value out of range
[E:][000030][CR][LF]
Invalid number of characters (between : and [CR][LF])
[E:][000012][CR][LF]
Setup
ZERO disabled
[E:][000060][CR][LF]
Device Status
Command not accepted due to local operation
[E:][000080][CR][LF]
Command not accepted due to synchronization, CLOSED or OPEN by digital input,
safety mode or fatal error
[E:][000082][CR][LF]
Hardware
Command not applicable for hardware configuration
[E:][000041][CR][LF]
Valve Cluster
Checksum failure (communication failure on cluster bus system)
[E:][000005][CR][LF]
Timeout expired (slave offline, no feedback from slave)
[E:][000006][CR][LF]
Control command rejected (slave valve not frozen)
[E:][000083][CR][LF]
Illegal slave command (e.g. sensor setup command)
[E:][000084][CR][LF]
Address out of range (slave address not in configured range)
[E:][000085][CR][LF]
Command not implemented on master valve (slave valve command only)
[E:][000086][CR][LF]
Command not implemented on slave valve (master valve command only)
[E:][000087][CR][LF]
Address information missing (G:xx prefix missing)
[E:][000088][CR][LF]